Portable Instruments
SIL_9_2011_MAS-100 NT Sampling head selection
We occasionally observe on customer instruments that the sampling head selected does not match the supplied head. Using a different sampling head then selected will lead to a deviation in sample volume.
SIL_7_2011_C&C SW 1.51 for MAS 100 NT_Ex
In the French language setting the user “calibreur” is not able to log in. The version 1.51
fixes this error.
There are no other changes to this version. All other languages are not affected and can work on
1.50 without restrictions.
SIL_6_2011_NiCd Battery Pack End of Life Notification
Cadmium is a toxic metal commonly found in some types of batteries like the ones used for the older version of MAS-100. Due to European law (Batterie Richtline 2006 / 66 / EG) this battery type is banned since September 26th 2010.
SIL_5_2011_MAS-100 NT New Mainboard (HW Rev. 4)
As part of the continuous improvement activities the new production batch of MAS-100 NT / Ex Mainboards have been slightly modified. These changes are reflected in the HW Rev. 4 coding. The Rev.2 and Rev. 3 mainboards are working fine and do not need to be changed.
None of the main functionality or the internal SW has been changed.
SIL_4_2011_MAS-100 NT Upgrade (HW, FW V1.4 and C&C V1.5)
This service information letter explains the upgrade of the MAS-100 NT.
SIL_3_2011_MAS-100 RCT Battery Problem
If the mainboard is on stock for more than 1-1.5 years the RCT battery can be completely
empty. This Service Information Letter explains the technical background of this effect and how it can be solved.
SIL_1_2011_MAS-100 CG® Ex Adjustment
Up to now the precision of the MAS-100 CG® Ex mass flow was specified with a symmetrical
tolerance band of +/- 5%.
The result interpretation of most of our customers however is asymmetric: a negative deviation (too little air sampled) is considered worse than a positive one (too much air sampled).
MBV AG decided to bring the adjustment limits in line with the result interpretation: the specification of the instrument precision will change to +10 / -0%.
SIL Service information Letter 11_2010
MBV AG has temporary stopped the delivery of MAS-100® NT and MAS-100® NT Ex due to technical problems.Please refer to the service information letter and the biological result interpretation from Merck.
SIL Service information Letter 10_2010
Customers observed a “Low Battery” message before reaching the specified sampling
volume of 18’000 L.
SIL Service information Letter 10_2010
MBV has introduced a new battery type for the MAS-100 Eco® (Ansmann+max1.2V/4500mAh). Due to the higher capacity the charging time changes the 9 hours
mentioned in the manual to 14-16 hours.
SIL Service information Letter 10_2010
Customers observed an alternation in air flow on a DA-100 when changing from the manualcalibration menu to the manual adjustment menu. The observed deviation is typically in the
range of 5%.
SIL Service information Letter 09_2010
MBV AG has changed the calibration equipment for the MAS-100 CG® Ex.SIL Service information Letter 09_2010
MBV AG has changed the calibration equipment for the MAS-100 CG® Ex.SIL Service Information Letter 04_2010
MAS-100® NT Exchange connector board.Please follow the instructions for changing the connector board.Use allways the new screws delivered with the new board.
SIL Service Information Letter 02_2010
New automated calibration procedure for MAS-100 CG® Ex starting 29. March 2010.Please note that the calibration certificate will change.




SIL_9_2011_MAS-100 NT Sampling head selection