FAQ MAS-100 Eco®

What is the main difference between the MAS-100Eco and the MAS-100?

The MAS-100Eco is especially designed for the food and environment industry. It has no integrated anemometer and no data communication possibilities.

Is it possible to use disinfectant solutions?

We do not recommend to use disinfectants because they may leave residues after application.

Is it possible to fix the MAS-100 to a tripod?

Yes. We supply a special adapter clamp.

Is it possible to use MAS-100Eco in the pharmaceutical industry?

We do not recommend the MAS-100Eco to use in the Pharmaceutical industry. The MAS-100 is especially designed for aerodynamics and very accurate running parameters.

How can I open the bottom plate of MAS-100Eco?

Take the perforated lid. It will fit into the small groves on the bottom plate and turn counter clock wise to open it.

What is a calibration reminder?

The calibration reminder is a soft ware function in the new MAS-100Eco air samplers which can be set after calibration between 0 and 12 month. After the set time has elapsed the display will show the message "Cal. now !" If calibration time is set to zero the reminder function is set off.

What can I do to change/stop the calibration reminder?

To stop the calibration reminder you have to put the value zero after the calibration step. To change the time for reminder choose between 1 and 12 month.