Breaking the seal will void the warranty regardless if it was broken by a service center or an end customer.
MBV may accept warranty claims with broken seals (for known issues) but these cases are decided by MBV on a case by case bases and are the sole decision of MBV.
Official Trainings and Training Certificate can only be issued by MBV. But if you want to involve your service engineers you can act as a supervisor:
you show them what they need to do and check their work / results. If everything is good you sign with a second signature and add your training certificate to the documentation package
The K3 value is a single reference point to adjust the Mass Flow sensor in the MAS-100 / MAS-100 ISO.
Calibration of the MAS-100™ (with a 400 x 0,7mm sampling head) flow rate involves adjustment of a single reference point, the K3 value. Following the calibration, a load adjustment is made to the fan motor, this is the K5 setting. The MAS-100 with firmware V2.6x and above allow for both the K3 and K5 value adjustments.
The K5 adjustment optimizes the voltage delivered to the motor, so that the MAS-100 maintains consistent flow rates with a flow restriction. A flow restriction is created to produce a load on the motor that far exceeds normal operating conditions.
The air flow adjustments made to the K5 setting do not affect the K3 calibration or the performance of the MAS- 100 under normal operating conditions (Sampling Mode). The flow rate measurement made during the K5 adjustment is not a direct indication of the performance of the mass flow sensor in the MAS-100. Therefore an MAS-100 should not be classified as out of tolerance if the SLPM measured in the K5 portion of the calibration protocol exceeds the optimum range previously specified.
| Following the calibration, a load adjustment is made to the fan motor, this is the K5 setting. This setting was necessary to compare the old readings on the MAS-100 C with the new resistance free anemometer. It is not needed anymore for the adjustment / calibration with a DA-100 /NT. Therefor it can be set to 0 at systems with a sampling head of 400 x 0.7 and it must be set to 0 at systems using a sampling head of 300 x 0.6, otherwise the fan would be loaded to high. MBV recommends to set the K5 value at all systems to 0. |
| There are no K3 and K5 settings in the NT systems. The most similar value is the PWM value. This sets the rotation speed of the fan motor. |
The impaction speed of the 400x0.7 sampling head is approx. 11m/s and with the 300x0.6 it is approx. 20m/s. Tests showed that with the 300x0.6 had been found more cfu counts than with the 400x0.7. The perforated sampling lid may be sterilized 20 Minutes at 121.
The inside of the sampling head may be cleaned under a laminar flow bench by spraying some alcohol on top of the perforated lid and leave the instrument running for about 5 minutes (500 Liters).
We do not recommend the use of disinfectant because they may leave residues after application. Some of them will have oxidizing agents which may interfere with the integrated anemometer.
Yes. Ask your local distributor. Please note: After update your MAS-100 must be calibrated!
Volumes set to zero will not be displayed anymore on display. In pharmaceutical companies the tested volume must always be 1000 Liters. If the other preprogrammed volumes are set to zero only 1000 Liter can be selected. No error in selecting volume is possible!
Any time. You set a zero volume to an other value and it will be selectable again in the standard menu.
MAS-100 may be upgraded for data communication starting with Serial No. 66161. Ask your local distributor for more information.
Portable instruments in general are not recommended to be used in Isolators. MAS-100ISO is especially designed to be used in isolators and clean rooms!
All MAS-100 instrument types comply with the: Biocontamination Control ISO/DIN 14698-1: Report - it is an impaction air sampler, - the impaction speed is > 20 m/s, - the isokinetic sampling speed is 045 m/s.
The MAS-100 has been tested with a particle counter type CLIMET CI-500: Result : >0,3µ:4 per cft / >0,5µ:2 per cft
- range calibration with DA-100 Anemometer possible = quicker and more accurate - for calibration it is not necessary any more to open MA-100 = easier handling - Delay function may be set on or off = easy to use - 5 volumes individually programmable = customer friendly - volumes set to zero will not be displayed= volume selection errors excluded - calibration reminder 0-12 month = convenience and security
Yes. The MAS-100 may by calibrated with the MAS-100C Calibration Unit or with the new DA-100 Anemometer. The MAS-100 is delivered with a calibration certificate.
We recommend to calibrate the MAS-100 once a year.
Yes. Since the MAS-100 has a built in anemometer the change of the perforated sampling head will not modify the calibration parameters.
The Anemometer is a device which measures the flow of a gas.
Yes. The anemometer which is built in to the MAS-100 will compensate for pressure and temperature variation.
The calibration reminder is a soft ware function in the new MAS-100 air samplers which can be set after calibration between 0 and 12 month. After the set time has elapsed the display will show the message "calibration time expired, please calibrate unit" If calibration time is set to zero the rem9inder function is set off.
Yes. Just continue pressing yes or no. The date of last calibration will be displayed and the the menu continues as usual.
To stop the calibration reminder you have to put the value zero after the calibration step. To change the time for reminder choose between 1 and 12 month.
Volumes set to zero will not be displayed anymore on display. In pharmaceutical companies the tested volume must always be 1000 Liters. If the other preprogrammed volumes are set to zero only 1000 Liter can be selected. No error in selecting volume is possible!
Push yes or no button. If nothing happens charge batteries for at least 3.5 hours. If still nothing happens on display, change battery pack.
Change battery pack.
Convince yourself.

It is absolutely no problem to calibrate old MAS-100 with DA-100. Just put the DA-100 on MAS-100 measure Volume and adjust K3 according to the deviation. Remark: If would like to compare the values obtained from DA-100 and MAS-100C (Calibration Unit) there might be a difference. That's why we have a update electronic chip V 2.6x to implement in old MAS-100. We have included K5 calibration factor , delay on/off, calibration reminder 5 individual programmable volumes, and for calibration the MAS-100 has not to be opened again.