See below the link to the current GWF accreditation certificate:
http://www.seco.admin.ch/sas_files/SCS-048-de.pdf
http://www.seco.admin.ch/sas_files/SCS-048-de.pdf
Please note that this figure is the worst case over the full measurement range. See the adjustment certificate for details (available on request).
What is the difference between the DA-100 Digital Anemometer and the MAS-100C Calibration unit?
Both instruments are made for flow calibration. The DA-100 is a turbine wing anemometer and is at least 20 times quicker than the MAS-100C Unit.
How do I calibrate the DA-100 Anemometer?
The DA-100 is delivered with a calibration certificate from a accredited Swiss federal laboratory. To recalibrate the DA-100 you may send it to a local laboratory or send it directly to the following address: MBV AG Service Department CH-6014 Littau Switzerland.
How frequently do I have to calibrate the DA-100?
We recommend to calibrate the DA-100 after approximately 2 years after the last calibration date.
The display on the DA-100 Anemometer is very weak
The battery is weak and has to be changed. The result displayed even on weak display is correct.
How can I change the battery in the DA-100 Anemometer?
Open the bottom plate with a perforated lid of an MAS-100 air sampler by turning counter clockwise. Remove old 9Volt battery and replace with a new one. Put the bottom plate back and screw it clock wise back . Do not remove any other parts from the DA-100.
How can I restart DA-100 if display is still on?
Just push the Start button for a few seconds and the DA-100 will switch to the waiting cycle and the start the measuring cycle.
Is it possible to calibrate MAS-100 with Serial. NO < 75000 with new DA-100 Anemometer?
It is absolutely no problem to calibrate old MAS-100 with DA-100. Just put the DA-100 on MAS-100 measure Volume and adjust K3 according to the deviation. Remark: If you compare the values obtained from DA-100 and MAS-100C (Calibration Unit) their might be a difference because we have changed the calibration procedure from point calibration to range calibration. Find more details in the DA-100 Manual. To correct the difference we supply an update electronic chip V 2.6x for the old MAS-100. Additional parameters are: K5 calibration factor for range calibration, delay on/off, calibration reminder, 5 individual programmable volumes and for calibration the MAS-100 has not to be opened again.